The outer shell is anti-static material, and a grounding wire to
ensure that valid and reliable measurement results, showing numbers
and graphics, and clear.
Can be measured with ion balance sheet, measuring ion generator
residual voltage can be measured range of + /-200V.
Field measurement range + /-20Kv * 2 LED determine the working
distance 25mm
Measurements can be equipped with ion balance board (+ /-200V)
Solely according to the zero reset
Solely according to the display lock
Idle Auto-Shutdown
Battery Performance Show
Function: Effective detection of ESD in the workplace equipment,
work surfaces, tools-of-band electrostatic charge.
Features: LCD / Level display test data, test data retention,
digital Clear display, high / low range.
Output: red LCD display positive charge, blue LCD display negative
charge measurement range: (0 ~200v low range), plus or minus
0.01kv-24kv (high range).
Test Accuracy10%
Power Supply: 9V
Accuracy: Low Range Accuracy: plus or minus 0.001kv; high range
accuracy: plus or minus 1kv
Input Voltage: 9V, 6F22 manganese battery
Electrostatic field can be measured :+/-( 0.00-20.0kV)
(automatically adjusted)
Standard Range: 25mm + /-1mm
Ion balance can be measured :+/-( 0-200V)
Measuring distance according to: ESD-STM3.1-2000 lonization
standard
C/RH:60%R LCD display update: 5 times per second reaction time: 1 second
working temperature and humidity :0-50 C / RH: 60% R
Graphic Display: Positive: red / minus: Blue Sound Alarm: On / idle
automatic shutdown / beyond the range can be measured automatically
shut down: idle for about 5 minutes to switch electrostatic field
measurement mode: Single-click Power button, there is a beep sound
of ion balance (Please Install ion balance measurement board) while
you press the Power and Zero, there is a sound beeps weight /
volume 0.17kg/12.3cm * 73cm * 2.5cm (with battery)
Key Note
1. Button "POWER" is the power switch.
2. Button "HOLD" is a digital lock.
3. Button "ZERO" is the number cleared.
4.Button "IB" is the unit switch.
Measurement Steps
1. By taking "POWER" button, so that the infrared lights, and then "ZERO" is cleared so that it displays zero; and then align objects being
tested; (At this point, infrared light to overlap together)
2. By taking "HOLD" key, lock the static object being tested voltage value, when the
recorded static pressure value, then press "HOLD" key to unlock.
When the need to measure other measured object, must click "ZERO" cleared to ensure accurate testing of static voltage.
Display
1. When the measuring instrument to the figures for the negative "-", then that is the measured object with a "negative voltage", when
the device is displayed as "+", then that is the measured object with a "positive voltage."
Notes
1. In the measurements, the need to connect "grounding line" in
order to avoid the measurement process, the measured object voltage
is too high, and thus burn out equipment.
2. When the measurement, when the instrument's "display" show the
number to 19000 volts or more (At this point, apparatus, will issue
a warning sound), this time we have to instantly removed the
equipment; possible causes are: ( a) Equipment burned. (B) electrostatic conductive device to the test through them too,
so that the human body with the ground instantaneous discharge (ie,
will electric shocks).